Important Control Solution Information
Use only UltraTRAK PRO control solutions.•
Check the expiration date on the control solution vial. •
Do not use if expired.
Control solution, meter, and test strips should come to •
room temperature 68°F to 77°F(20°C to 25°C) before
testing.
Shake the vial, discard the rst drop of control solution, •
and wipe off the dispenser tip to ensure a good sample
and an accurate result.
Use only for 90 days after rst opening. Record the •
discard date (date opened plus 90 days) on the control
solution vial. Discard after 90 days.
Store the control solution tightly closed at temperatures •
36°F to 86°F (2°C to 30°C). Do not freeze.
The control solution range printed on the test strip vial is for UltraTRAK
PRO control solution only. It is used to test meter and test strip perfor-
mance. It is not recommended range for your blood glucose level.
PLEASE NOTE!