Vertex Diagnostics UltraTRAK Pro Manual do Proprietário Página 45

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 68
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 44
www.vertexdiagnostics.com
vertex
diagnostics
45
You may still have a variation from the result because
blood glucose levels can change signicantly over short
periods, especially if you have recently eaten, exercised,
taken medication, or experienced stress*
2
. In addition, if
you have eaten recently, the blood glucose level from a
nger stick can be up to 70 mg/dL (3.9 mmol/L) higher than
blood drawn from a vein (venous sample) used for a lab
test*
3
. Therefore, it is best to fast for eight hours before
doing comparison tests. Factors such as the amount of red
blood cells in the blood (a high or low hematocrit) or the
loss of body uid (severe dehydration) may also cause a
meter result to be different from a laboratory result.
References
*2: Surwit, R.S., and Feinglos, M.N.: Diabetes Forecast (1988), April,
49-51.
*3: Sacks, D.B.: “Carbohydrates. “ Burtis, C.A., and Ashwood, E.R.( ed.),
Tietz Textbook of Clinical Chemistry. Philadelphia: W.B. Saunders Com
pany (1994), 959.
Vista de página 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 67 68

Comentários a estes Manuais

Sem comentários